ACR Systems SRP-001-128K

SmartReader Plus 1 2-Channel (Temperature and External Temperature/Resistance/Switch) Data Logger with 128 KB Memory

SRP001128K
SRP-001-128K
01-0112

ACR Systems SRP-001-128K

SmartReader Plus 1 2-Channel (Temperature and External Temperature/Resistance/Switch) Data Logger with 128 KB Memory

SRP001128K
SRP-001-128K
01-0112

The ACR Systems SRP-001-128K SmartReader Plus 1 is a 2-Channel (Temperature and External Temperature/Resistance/Switch) Data Logger. It has an internal ambient temperature channel and input for optional remote temperature or contact closure sensing. Logger setup is easy with built-in menus for sample rate and mode, start delay, and equations.

  • Operate with TrendReader 2 Software
  • Number of Channels: 2
  • Memory Size: 128 KB
  • Connect up to 10 data loggers
$827.00
This item ships FREE
  • Note: SmartReader Data Loggers require TrendReader 2 Software to operate
  • Note: Probes, transformers and transducers not included
  • Number of Channels: Two (2)
    • One (1) channel for ambient temperature
    • One (1) channel for temperature (with thermistor probe), switch status or resistance
  • Monitor and record temperature
  • Option to monitor remote temperature, resistance or switch status
  • Use with variety of different temperature probes:
    • General purpose
    • High temperature
    • Low temperature
    • Waterproof temperature
    • Oven temperature
    • Skin surface temperature
    • Pipe surface temperature
  • Memory Size: 128 KB
  • 12-bit resolution
  • 10-year battery life
  • 3-year warranty
  • Fast sample rates
  • Remote access via modem or Ethernet
  • Network capable (connect up to 10 data loggers)
  • Alarm dial-out
  • User configurable
  • Case Material: Noryl plastic
  • External Connector: Removable screw-type terminal strip with Common (-) connection
  • Mounting: Magnetic backing or locking eyelet
  • Operating Limits: -40 to 70°C (-40 to 158°F) and 0 to 95% relative humidity (non-condensing)
  • Clock Accuracy: +/- 2 seconds per day
  • Battery: 3.6 volt Lithium, 1 Amp-Hour
  • Battery Life: 10 years under normal use (factory replaceable)
  • Power Consumption: 5 to 10 microamps (continuous)
  • Sampling Methods:
    • Continuous (First-in, First-out - not available with sample rates faster than 8 seconds)
    • Stop when full (Fill-then-stop)
  • Sampling Rates: User selectable rates from 25 per second (0.04 seconds) to once every 8 hrs
  • Resolution: 12-bit (1 part in 4,096)

Internal Temperature Sensor

  • Type: Negative Temperature Coefficient Thermistor 10,000 ohms @ 25°C (77°F)
  • Range: -40 to 70°C (-40 to 158°F)
  • Accuracy: +/- 0.2°C over the range of 0 to 70°C (+/- 0.3°F over the range of 32 to 158°F)
  • Resolution: 0.03°C at 25°C (0.05°F at 77°F); better than 0.07°C (0.12°F) between -25 and 70°C (-13 and 158°F); better than 0.13°C (0.23°F) between -40 and -25°C (-40 and -13°F)

Applications

  • Food process verification
  • Pharmaceutical and/or laboratory storage
  • Transportation of temperature-sensitive goods
  • Equipment run time
  • HVAC system testing and balancing

The ACR Systems SmartReader Plus 1 (SRP-001-128K) 2-Channel (Temperature and External Temperature/Resistance/Switch) Data Logger with 128 KB Memory is specifically designed for easy monitoring and recording of temperature. With one internal ambient temperature channel and one input for optional remote temperature or contact closure sensing, this data logger is ideal for quick and accurate temperature recordings in a variety of applications.

TrendReader 2 Software

The ACR Systems TrendReader 2 Software is a powerful, versatile and easy-to-use application, designed exclusively for ACR's SmartReader Plus, SmartReader, IAQ, OWL, TRH, JR and Nautilus data loggers. Powerful features and easy setup enables collected data to be thoroughly analyzed in seconds.

Downloading and Communicating with Data Logger

Setup, download and view information from a single data logger with ACR's IC-102 interface cable (included in software interface package). Connection is simple; plug the interface cable into the USB port of your computer and connect the other end to the logger. No tools, cards or docking stations are required.

Data Logger Setup

With built-in menus for sample rate and mode, start delay, equations and more, logger setup is fast and easy. In addition, equations are already included for all ACR sensors so you need only select the appropriate equation from the menu or customize your own. Real-time readings are displayed in the Setup window, allowing you to verify that your data logger is working correctly before placing it in the field.

Viewing Data

With TrendReader 2, data can be viewed in several ways. For presentation purposes, view your collected data in graphical format or for a more detailed analysis, view your data in table format. Either way, multiple channels of data can be displayed at once. Multiple graphs can be displayed together and compounded to create new graphs. For added presentation power, add comments, change graph colors, zoom in on specific portions of data, and display your data using custom engineering units. Print or copy and paste graphs into your reports and presentations. Select all, or portions of the table and copy and paste into your favorite Windows program.

Exporting Data

Use the export function in TrendReader 2 to export your data to other programs for further analysis. Data can be exported in TXT, BMP, PDF, JPG or CSV formats.

More Information
BrandACR
Catalog01-0112
ConditionNEW
Vendor Alias01-0112