ACR Systems SRP-006-128K

SmartReader Plus 6 8-Channel (Temperature/Cold-Junction Compensation and External Temperature) Data Logger with 128 KB Memory

SRP006128K
SRP-006-128K
01-0122

ACR Systems SRP-006-128K

SmartReader Plus 6 8-Channel (Temperature/Cold-Junction Compensation and External Temperature) Data Logger with 128 KB Memory

SRP006128K
SRP-006-128K
01-0122

The ACR Systems SRP-006-128K SmartReader Plus 6 is an 8-Channel (Temperature/Cold-Junction Compensation and External Temperature) Data Logger. Easily record temperature from any J, K, S, T or E type thermocouple probe. With 7 inputs for thermocouple probes and an on-board cold-junction thermistor, it is one of the most versatile thermocouple data loggers available.

  • Operate with TrendReader 2 Software
  • Memory Size: 128 KB
  • Connect up to 10 data loggers
$1,472.00
This item ships FREE
  • Note: SmartReader Data Loggers require TrendReader 2 Software to operate
  • Note: Probes, transformers and transducers not included
  • Number of Channels: Eight (8)
    • One (1) channel for ambient temperature (also for coldjunction compensation)
    • Seven (7) channels for temperature (with thermocouples)
  • Monitor and record temperature in eight different locations
  • Monitor extremely high temperatures
  • Monitor temperature over a wide range
  • Use with variety of different temperature probes:
    • General purpose
    • High temperature
    • Low temperature
    • Waterproof temperature
    • Oven temperature
    • Skin surface temperature
    • Pipe surface temperature
  • Greater accuracy and resolution than SR6
  • Memory Size: 128 KB
  • 12-bit resolution
  • 10-year battery life
  • 3-year warranty
  • Fast sample rates
  • Remote access via modem or Ethernet
  • Network capable (connect up to 10 data loggers)
  • Alarm dial-out
  • User configurable
  • Case Material: Noryl plastic
  • External Connector: Removable screw-type terminal strip with Common (-) connection
  • Mounting: Magnetic backing or locking eyelet
  • Operating Limits: -40 to 70°C (-40 to 158°F) and 0 to 95% relative humidity (non-condensing)
  • Clock Accuracy: +/- 2 seconds per day
  • Battery: 3.6 volt Lithium, 1 Amp-Hour
  • Battery Life: 10 years under normal use (factory replaceable)
  • Power Consumption: 5 to 10 microamps (continuous)
  • Sampling Methods:
    • Continuous (First-in, First-out - not available with sample rates faster than 8 seconds)
    • Stop when full (Fill-then-stop)
  • Sampling Rates: User selectable rates from 25 per second (0.04 seconds) to once every 8 hrs
  • Resolution: 12-bit (1 part in 4,096)

Internal Temperature Sensor

  • Type: Negative Temperature Coefficient Thermistor 10,000 ohms @ 25°C (77°F)
  • Range: -40 to 70°C (-40 to 158°F)
  • Accuracy: +/- 0.2°C over the range of 0 to 70°C (+/- 0.3°F over the range of 32 to 158°F)
  • Resolution: 0.03°C at 25°C (0.05°F at 77°F); better than 0.07°C (0.12°F) between -25 and 70°C (-13 and 158°F); better than 0.13°C (0.23°F) between -40 and -25°C (-40 and -13°F)
  • Accuracy: +/- 0.5% of range + resolution

Applications

  • Monitor performance of gas-fired heating equipment
  • Verifying concrete curing temperatures
  • Process verification

The ACR Systems SmartReader Plus 6 (SRP-006-128K) 8-Channel (Temperature/Cold-Junction Compensation and External Temperature) Data Logger with 128 KB Memory allows you to monitor temperature over a wide range. Easily and accurately record temperature from any J, K, S, T or E type thermocouple probe. At 12-bit resolution, the SmartReader Plus 6 offers greater accuracy and more precise measurements than the SmartReader 6. With seven inputs for thermocouple probes and an on-board cold-junction thermistor, this logger is one of the most versatile thermocouple data loggers available.

TrendReader 2 Software

The ACR Systems TrendReader 2 Software is a powerful, versatile and easy-to-use application, designed exclusively for ACR's SmartReader Plus, SmartReader, IAQ, OWL, TRH, JR and Nautilus data loggers. Powerful features and easy setup enables collected data to be thoroughly analyzed in seconds.

Downloading and Communicating with Data Logger

Setup, download and view information from a single data logger with ACR's IC-102 interface cable (included in software interface package). Connection is simple; plug the interface cable into the USB port of your computer and connect the other end to the logger. No tools, cards or docking stations are required.

Data Logger Setup

With built-in menus for sample rate and mode, start delay, equations and more, logger setup is fast and easy. In addition, equations are already included for all ACR sensors so you need only select the appropriate equation from the menu or customize your own. Real-time readings are displayed in the Setup window, allowing you to verify that your data logger is working correctly before placing it in the field.

Viewing Data

With TrendReader 2, data can be viewed in several ways. For presentation purposes, view your collected data in graphical format or for a more detailed analysis, view your data in table format. Either way, multiple channels of data can be displayed at once. Multiple graphs can be displayed together and compounded to create new graphs. For added presentation power, add comments, change graph colors, zoom in on specific portions of data, and display your data using custom engineering units. Print or copy and paste graphs into your reports and presentations. Select all, or portions of the table and copy and paste into your favorite Windows program.

Exporting Data

Use the export function in TrendReader 2 to export your data to other programs for further analysis. Data can be exported in TXT, BMP, PDF, JPG or CSV formats.

More Information
BrandACR
Catalog01-0122
ConditionNEW
Vendor Alias01-0122