Fluke Calibration 5615-6-P

Secondary Standard PRT Probe with INFO-CON for 1523/1524, 6 in. x .188 in., -200 °C to 300 °C, 6 ft. Lead

56156P
5615-6-P

Fluke Calibration 5615-6-P

Secondary Standard PRT Probe with INFO-CON for 1523/1524, 6 in. x .188 in., -200 °C to 300 °C, 6 ft. Lead

56156P
5615-6-P
  • Termination: INFO-CON for Models 1523 and 1524
  • Temperature range: -200 °C to 300 °C
  • Nominal resistance at 0 °C: 100 Ohm ± 0.10 Ohm
This product is available for shipping to the US, Canada, and Puerto Rico only.
$1,555.95 Price $1,638.00
This item ships FREE
  • Termination: INFO-CON for Models 1523 and 1524
  • Temperature range: -200 °C to 300 °C
  • Nominal resistance at 0 °C: 100 Ohm ± 0.10 Ohm
  • Temperature coefficient 0.0039250 Ohm/Ohm/°C
  • Accuracy (includes calibration and 100 hr drift (k=2)): ±0.024 °C at -200 °C; ±0.012 °C at 0 °C; ±0.035 °C at 420 °CC
  • Short-term repeatability (three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)): ±0.009 °C at 0.010 °C
  • Drift (After 100 hrs at max temp, 95 % confidence (k=2)): ±0.007 °C at 0.010 °C
  • Sensor length: 28 mm (1.1 in.)
  • Sensor location: 6.9 mm ± 3.3 mm from tip (0.27 in. ± 0.13 in.)
  • Sheath dimensions: 152 mm x 4.76 mm (6.0 in. x 0.188 in.)
  • Sheath diameter tolerance: ±0.127 mm (±0.005 in.)
  • Sheath material: Inconel 600
  • Minimum insulation resistance: 1000 MOhm at 23 °C
  • Transition junction temperature range (temperatures outside this range will cause irreparable damage; for best performance, transition junction should not be too hot to touch): -50 °C to 200 °C
  • Transition junction dimensions: 71 mm x 13 mm dia (2.8 in. x 0.5 in.)
  • Maximum immersion length: 102 mm (4 in.)
  • Response time (Per ASTM E 644): 9 seconds typical
  • Self heating (in 0 °C bath): 50 mW/°C
  • Lead-wire cable type: PTFE insulated with PTFE jacket, 22 AWG
  • Lead-wire length: 183 cm (72 in. or 6 ft.)
  • Lead-wire temperature range: -50 °C to 200 °C
  • Calibration: NVLAP-accredited calibration included, lab code 200706-0

The Fluke Calibration 56156P Secondary Standard PRT Probe with INFO-CON for 1523/1524, 6 in. x .188 in., -200 °C to 300 °C, 6 ft. Lead has an Inconel 600 sheath that is 152 mm (6 in.) long and 4.76 mm (0.188 in.) in diameter. It is a secondary reference temperature standard designed to bridge the gap between the highest laboratory standards and industrial or second-tier lab locations. It has accuracy of ± 0.012 °C at 0.01 °C.

The element is constructed of reference-grade platinum wire (99.999% pure) for excellent stability. The wire is wound into a coil and placed in a mandrel where it is uniformly supported in a manner that virtually eliminates hysteresis. The electrical configuration is a four-wire current-potential hookup that eliminate the effects of lead-wire resistance.

This Inconel-sheathed probes has a fully supported sensing element, making it more durable than an SPRT. The element is protected in an ultra-high-purity ceramic case with a hermetic glass seal to improve output stability by locking out moisture and contaminants.

The Fluke Calibration 5615-6-P comes calibrated with ITS-90 coefficients, making it compatible with many excellent readout devices, including Fluke Calibration's 1529 Chub-E4, 1560 Black Stack, and 1502A Tweener. It bridges the gap between a 100-ohm industrial RTD and an SPRT.

More Information
Country of OriginUnited States
BrandFluke Calibration
ConditionNEW
  • 1 - Coefficient Sheet
  • 1 - User Manual CD
  • 1 - Certification Sheets