Fluke Calibration 5627A-6-J
Precision Industrial RTD Probe with Banana Plugs, 6 in. x 3/16 in., -200 °C to 300 °C
Fluke Calibration 5627A-6-J
Precision Industrial RTD Probe with Banana Plugs, 6 in. x 3/16 in., -200 °C to 300 °C
- Brand Name: Fluke Calibration
- Condition: NEW
- Country of Origin: United States
- Termination: Banana plugs
- Resistance: Nominal 100 Ohm
- Temperature coefficient: 0.00385 Ohm/Ohm/°C nominal
- Temperature range: -200 °C to 300 °C (transition and cable temperature: 0 °C to 150 °C)
- Drift rate (k=2): ±0.04 °C at 0 °C after 100 hours at 420 °C
- Sheath material: 316 stainless steel
- Leads: PTFE-insulated, nickel-plated stranded copper, 22 AWG
- Time constant: Four seconds maximum for 63.2% response to step change in water moving at 3 fps
- Bending radius: Sheath may be ordered with a bend on minimum radius of 19 mm (3/4 in.) except for 50 mm (2 in.) area of sheath near tip
- Calibration: Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs; T values in 1 °C increments from -196 °C to 300 °C; ITS-90 coefficients included
- Immersion: At least 100 mm (4 in.) recommended
- Calibrated accuracy (includes calibration uncertainty and 100 hr drift) (k=2): ±0.026 °C at -196 °C; ± 0.046 °C at 0 °C; ±0.077 °C at 200 °C; ±0.124 °C at 420 °C
- Size: 152 mm x 4.7 mm (6 in. x 3/16 in.)
The Fluke Calibration 5627A6J Precision Industrial RTD Probe with Banana Plugs, 6 in. x 3/16 in., -200 °C to 300 °C has a temperature range of up to 300 °C and an accuracy as good as ± 0.025 °C. This precision thermometer is shipped with its ITS-90 coefficients and a calibration table in 1 °C increments. One of the best features of the 5627A-6-J sensor is that it conforms to the standard 385 curve, letting you use your DIN/IEC RTD meters fully.
The 5627A-6-J is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. It is individually calibrated and includes a NVLAP-accredited report of calibration from the manufacturer with lab code 200706-0.
Country of Origin | United States |
---|---|
Brand | Fluke Calibration |
Condition | NEW |
- 1 - Coefficient Sheet
- 1 - User Manual CD
- 1 - Certification Sheets