Lascar Electronics EL-WIFI-DULT

WiFi Dual Channel Ultra Low Temperature Cryogenic Vaccine Data Logger with Display, Cloud-Enabled, -148 to 212°F

ELWIFIDULT
EL-WIFI-DULT

Lascar Electronics EL-WIFI-DULT

WiFi Dual Channel Ultra Low Temperature Cryogenic Vaccine Data Logger with Display, Cloud-Enabled, -148 to 212°F

ELWIFIDULT
EL-WIFI-DULT

The Lascar Electronics EL-WIFI-DULT WiFi Dual Channel Ultra Low Temperature Cryogenic Vaccine Data Logger with Display is designed to monitor vaccines in cryogenic dry ice storage and transit.

  • Temperature measurement range: -148 to 212°F (-100 to 100°C)
  • Designed to monitor vaccines in cryogenic dry ice storage and transit
  • Performs dual channel recording through two external T-type thermocouple probes
$257.30 Price $390.49
This item ships FREE
In stock: 1 Units
  • Temperature measurement range: -148 to 212°F (-100 to 100°C)
  • Designed to monitor vaccines in cryogenic dry ice storage and transit
  • Performs dual channel recording through two external T-type thermocouple probes
  • Wirelessly stream and view data on the EasyLog Cloud, App, or on a PC
  • View and analyze multiple channels, including graphing of historic data
  • Configurable high and low alarms with on-screen indicator
  • Alarm functions include email and SMS notifications
  • Stores data internally if WiFi connection is interrupted, automatically uploading data to Cloud once reconnected
  • Wall mountable with included bracket
  • Easy to set up using free PC software or mobile app

Specifications

  • Temperature measurement range: -148 to +212°F (-100 to +100°C)
  • Temperature measurement resolution: 0.2°F (0.1°C)
  • Accuracy: ±4.5°F (±2.5°C) typical
  • Logging period: User selectable between 10 seconds and 12 hours
  • Transmission period: User selectable between 1 min and 24 hours
  • Operating temperature range (logger): -4 to +140°F (-20 to +60°C)
  • Battery life: >6 months typical
  • Dimensions: 3.22 x 2.75 x 1.41 in. (82 x 70 x 36 mm)
  • Probe dimensions: 11.8 x Φ 0.05 in., 39.3 in. cable (300 x Φ 1.5 mm, 1m cable)
  • IP Rating: IP40

The Lascar Electronics EL-WIFI-DULT WiFi Dual Channel Ultra Low Temperature Cryogenic Vaccine Data Logger with Display is designed to monitor vaccines in cryogenic dry ice storage and transit. The measurement range is -100 to +100°C (-148 to +212°F) using the probes supplied with the logger. The logger automatically records data and uploads it to the EasyLog Cloud via a standard wireless network; the data can be viewed on a PC using free software or the EasyLog Cloud, or on a mobile device using the EasyLog Cloud App.

To configure the EL-WIFI-DULT data logger for use on a given wireless network, either connect it via USB to a PC running EasyLog WiFi software on the network, or configure wirelessly using the EasyLog Cloud app on a mobile phone with access to the network. The logger can then be placed anywhere within range of the network. If the logger temporarily loses connectivity with the network, it will log readings until it is able to communicate again with the PC application or EasyLog Cloud (max 30 days at 10 second sample interval).

The device is IEEE 802.11bgn (2.4GHz) compliant, supports WEP, WPA/WPA2 encryption and enterprise networks (PEAP, TTLS, FAST).

The EL-WiFi-DULT has a protection rating of IP40. The unit is freestanding, but can be attached to a wall or surface using the included bracket. The unit can be clipped in and out of the bracket as required.

More Information
BrandLascar
ConditionNEW
Vendor AliasEL-WIFI-DULT
  • 1 - Wireless 2-Channel Ultra Low Temperature Data Logger (EL-WIFI-DULT)
  • 2 - T-Type Thermocouple Probes with 1m Cable (EL-P-TC-T-ULT)
  • 1 - Wall Mounting Bracket (EL-WIFI WALL BRACKET)
  • 1 - USB Type A to Micro B Cable (CABLE USB A-MICRO B)