- Brand Name: Fluke Calibration
- Condition: NEW
- Country of Origin: Japan
- Termination: INFO-CON for Models 1523 and 1524
- Temperature range: -200 °C to 350 °C
- Nominal RTPW: 100 Ohm
- Temperature coefficient: 0.00385 Ohm/Ohm/°C nominal
- Uncalibrated DIN/IEC conformity: DIN/IEC Class A; ±0.15 °C at 0 °C
- Thermal EMF: 20 mV at 350 °C
- Cable: PVC, 4-wire cable, 2 meters long, 90 °C max temp
- Calibrated probe accuracy (includes calibration uncertainty and short-term stability): ±0.04 °C at -200 °C; ±0.04 °C at 0 °C; ±0.045 °C at 200 °C; ±0.055 °C at 300 °C
- Time constant (63.2 %) from 0 °C to 100 °C: 3.0 seconds
- Immersion depth: 32 mm (1.25 in.)
- Sheath 316 SST: 200 x 1.6 mm (8 x 0.06 in.)
- Designed for temperature measurements requiring fast response or short immersion over wide range
The Fluke Calibration 562216P Fast Response PRT Probe with INFO-CON for 1523/1524, 1.6mm (0.06 in.), 6 ft. Lead features a stainless steel sheath. And because this high-quality wire-wound sensor comes in a small package, heat transfer to the sensors occurs quickly.
Immersion requirements for this probe are also a plus, with an immersion depth of 32 mm (1.25 in.). Getting into shallow or tight places is not a problem, either. And because this probe can handle temperatures from -200 °C to 350 °C, it is more versatile than most thermistors. Uncalibrated, this probe conforms to DIN/IEC Class A requirements with accuracy of ±0.15 °C at 0 °C.
Readout options for the Fluke Calibration 5622-16-P includes the 1523/1524 Handheld Thermometers, or the 1529 Benchtop Thermometer. Each of these readouts will read your PRT as a standard DIN/IEC probe or as an individually calibrated PRT.
Country of Origin | Japan |
---|---|
Brand | Fluke Calibration |
Condition | NEW |