Your company account is blocked and you cannot place orders. If you have questions, please contact your company administrator.

A&D Weighing RA30-112

Remote Control Module, 30V Maximum

RA30112
RA30-112

A&D Weighing RA30-112

Remote Control Module, 30V Maximum

RA30112
RA30-112

The A&D Weighing RA30-112 Remote Control Module is for use with the Omniace RA3100 Data Acquisition System. It includes three major types of functions: the remote function, carrier function for the AC strain module, and waveform confirmation output for the voltage input module.

  • Start/stop, mark printing, and paper feed
  • External sampling by external signal
  • External trigger signal input and output
  • Bridge power supply to 2ch AC strain module
$540.00
This item ships FREE

Key Specifications

  • Input Connector: half-pitch 20-pin connector
  • Output Connector: half-pitch 14-pin connector
  • External Output:
    • Control Signal: START/STOP, MARK, FEED, PRINT, TRIG
    • Input Level: High level: 2.1V to 5.0V, Low level: 0V to 0.5V (active low)
    • Response Speed: Select from High-speed/Normal/Low-speed
    • Effective Pulse Width:
      • High-Speed Response: 1μs or higher during high interval, 1μs or higher during low interval
      • Normal Response: High interval 1ms or higher, Low interval 1ms or higher
      • Low-Speed Response : 10ms or higher during high, 10ms or higher during low-speed response
    • Max. Allowable Input Voltage: 30V
  • External Output: Function: Control signals can be externally output
    • Control Signal: START/STOP, MARK, FEED, PRINT, TRIG, EXT1/EXT2
    • Output Level: High level: 3.8V to 5.0V, Low level: 0V to 0.5V (active low)
    • Output Pulse Width:
      • START/STOP, FEED, PRINT: Active output during operation
      • TRIG, MARK: High-speed response: 1μs/Normal response : 1ms
      • Low-speed response: 10ms
  • External Sampling Input (EXT.SMPL IN): Synchronization via external clock signal is possible (simultaneous SSD recording and printer recording are not possible)
    • Input Level: High level: 2.1V to 5.0V, Low level : 0V to 0.5V
    • Effective Pulse Width: High-speed (SSD Recording): 2μs or higher/Low-speed (Printer Recording): 1ms or higher
    • Maximum Input Frequency: High-speed (SSD Recording): 250kHz/Low-speed (printer recording): 500Hz, 0.1mm/pulse
  • External Sampling Output (EXT.SMPL OUT): Function: Synchronization clock signal can be output externally
    • Output Level High Level: 3.8V to 5.0V, Low level: 0V to 0.5V (active low)
  • Synchronization Signal for AC Strain Input/Output Function: Synchronization signal generator for using AC strain
    • Carrier Wave: 0V to 5V, square wave, 5kHz
    • Synchronization: Synchronization possible with other RA3000 product including RA30-112
  • Reference Clock for Calibration: Square wave signal output for operation check of voltage input module
    • Output Level: 0V to 5V (±1%)
    • Output Frequency: 1kHz (±1%)
    • Duty Ratio: 50% (±5%)
  • Withstand Voltage: AC300V, 1 minute (between input/output and main chassis)
  • Maximum Rated Voltage to Ground: AC, DC 42V
  • Dimensions: Approx. 140 (input-side W) x 223(D) x 20(H) mm
  • Weight: Approx. 250g
  • Compliance Standards:
    • Safety: EN61010-1
    • EMC: EN61326-1, Class A
Model No.
Qty.
Cart Subtotal (%1 items):
All taxes, shipping fees and discounts will be applied at checkout.